Alco-Screen .02 Alcohol Test Procedure

Abstain from placing anything in the subjects mouth for fifteen (15) minutes
prior to beginning the test. This includes non-alcoholic drinks,
tobacco products, coffee, breath mints, food, etc.

Open the foil
package and remove the test strip. The preservative packet is
not needed to perform the test and should be discarded. Observe
the reactive pad on the end of the test strip. The pad should be
a light cream color. A test strip with a reagent pad which has a
colored line or is otherwise discolored must be discarded.

Saturate the reactive pad with saliva from a sputum cup or by
applying saliva directly to the pad. Immediately start timer.
After ten seconds, shake off excess saliva.

At four (4) minutes,
observe the results on the reactive pad. Reading the results may
be made easier by placing the test on a white background. The
development of a distinct colored line across the reactive pad
indicates the presence of alcohol greater than 0.02% (0,2) BAC.
Results obtained after more than five (5) minutes may be
erroneous.
Positive Result: Blood alcohol above 0.02%

Negative: Blood alcohol less
than 0.02%.

"Legal limit
varies according to testing location".
|